Ralston Instruments FLP1-GJ-2M FieldLab Pressure Calibrator
Request a quote
Ralston Instruments FLP1-GJ-2M FieldLab is a Class 1 Division 1 intrinsically safe portable pressure calibrator with 100 PSI range, configurable test modes, and 2 million data points storage.
Product details
| Model | FLP1-GJ-2M |
| Manufacturer | Ralston Instruments |
| Category | General Pressure Calibrators |
| Availability | Made to order |
Description
Overview
Ralston Instruments FLP1-GJ-2M FieldLab is a portable pressure calibrator rated for 100 PSI (7 bar / 700 kPa). Designed for hazardous area use, it holds Class 1 Division 1 intrinsically safe certification with ATEX/IECEx Zone 0 approval. The device features desktop PC software for custom test modes, automated data logging, and traceable calibration management.
Key Features
- 100 PSI (7 bar / 700 kPa) maximum pressure range
- Class 1 Division 1 intrinsically safe with ATEX/IECEx Zone 0 and CE approval
- Up to 2 million data points onboard storage
- Configurable Test Modes for repeatability and traceability
- 3.5-inch color TFT LCD display
- IP67 rated for wet environments
- Lithium-ion rechargeable battery (2-month runtime)
- Free FieldLab Desktop Software for test configuration and data management
Included Accessories
- NIST traceable calibration certificate
- 2-meter shielded USB Type B cable
- Power adapter with international plugs (US, Europe, Australia, China, UK)
Applications
- Pressure gauge and transmitter calibration in hazardous areas
- Safety valve testing and validation
- Field calibration in ATEX/IECEx classified zones
- Compliance and regulatory testing
Specifications
| Pressure Range | 0 to 100 psi (0 to 7 bar, 0 to 700 kPa) |
| Accuracy (20-100% of Full Scale) | ±0.1% of reading |
| Accuracy (0-20% of Full Scale) | ±0.02% of full scale |
| Vacuum Accuracy | ±0.25% of full scale (ranges 300 psi and lower) |
| Operating Temperature | -10 to 50°C (14 to 122°F) |
| Storage Temperature | -40 to 75°C (-40 to 167°F) |
| Display | 3.5-inch color TFT LCD |
| Battery | Lithium-ion rechargeable, 4.8 Ah capacity, 2-month runtime |
| IP Rating | IP67 (1 meter for 30 minutes) |
| Wetted Material | 316 stainless steel |
| Data Storage | Up to 2 million data points |
| Communication | USB Type B |
| Connection | 1/4 inch male NPT |
| Certification | Class 1 Division 1 intrinsically safe, ATEX/IECEx Zone 0, CE approved |
Request a quote
Ralston Instruments FLP1-GJ-2M FieldLab Pressure Calibrator
Response within 24 hours
No obligation
Direct communication