Keysight 1146A AC/DC current probe measures currents from 100 mA to 100 A rms using Hall-effect technology, DC to 100 kHz, without breaking the circuit.
Product details
| Model | 1146A |
| Manufacturer | Keysight |
| Category | Oscilloscope Probes |
| Availability | Made to order |
Description
Overview
The Keysight 1146A is an AC/DC current probe that provides accurate measurement of currents from 100 mA to 100 A rms without breaking the circuit. Using Hall-effect sensor technology, it can measure both AC and DC signals from DC to 100 kHz. The probe is compatible with all Keysight oscilloscopes and any scope with 0.2 to 0.5 V/div sensitivity, a minimum input impedance of 1 MΩ, and a BNC input connector.
Key Features
- Frequency range: DC to 100 kHz (-3dB with current derating)
- Current measurement ranges: 100 mV/A (100 mA to 10 A peak) or 10 mV/A (1 to 100 A peak)
- Hall-effect sensor technology for AC and DC measurements
- Maximum working voltage: 600 V
- Battery: 9 V alkaline
- Non-invasive measurement without breaking the circuit
Compatibility
- All Keysight oscilloscopes
- Any oscilloscope with 0.2 to 0.5 V/div sensitivity
- Minimum input impedance: 1 MΩ
- BNC input connector
- Compatible with Keysight 1000, 3000, 5000, 6000, 7000, 8000, 9000, 80000, and 90000 Series
Applications
- Power supply and converter measurements
- Motor controller current monitoring
- Switching circuit analysis
- Power distribution diagnostics
Specifications
| Frequency Range | DC to 100 kHz (-3dB) |
| Current Range (100 mV/A) | 100 mA to 10 A peak |
| Current Range (10 mV/A) | 1 to 100 A peak |
| AC Current Accuracy (100 mV/A) | 3% of reading ≥ 50 mA |
| AC Current Accuracy (10 mV/A, 500 mA to 40 A) | 4% of reading ≥ 50 mA |
| AC Current Accuracy (10 mV/A, 40 to 100 A) | 15% maximum at 100 A |
| Working Voltage | 600 V max |
| Battery | 9 V alkaline |
| Sensor Technology | Hall-effect |
| Input Impedance | ≥ 1 MΩ |