Fluke 9144-B-P-156 Field Metrology Well
Request a quote
Fluke 9144-B-P-156 field metrology well with process electronics, 50–660 °C range, ±0.35 °C accuracy, and built-in thermometer for RTD and 4-20 mA calibration.
Product details
| Model | 9144-B-P |
| Manufacturer | Fluke Calibration |
| Category | Dry Block Calibrators |
| Availability | Made to order |
Description
Overview
The Fluke 9144-B-P-156 field metrology well extends high-performance temperature calibration to industrial process environments with integrated process electronics. It covers 50 °C to 660 °C and includes a built-in two-channel thermometer for measuring PRTs, RTDs, thermocouples, and 4-20 mA transmitters with 24 V loop power. Automated test documentation up to 20 tests with unique alphanumeric IDs makes this a complete field calibration solution.
Key Features
- Temperature Range: 50 °C to 660 °C (122 °F to 1220 °F)
- Display Accuracy: ±0.35 °C at 50 °C, ±0.35 °C at 420 °C, ±0.5 °C at 660 °C
- Stability: ±0.03 °C at 50 °C, ±0.04 °C at 420 °C, ±0.05 °C at 660 °C
- Built-in two-channel thermometer for PRT, RTD, thermocouple, and 4-20 mA measurements
- 24 V loop power supply for transmitter operation
- Onboard non-volatile memory for documenting up to 20 tests
- Radial Uniformity: ±0.02 °C at 50 °C, ±0.05 °C at 420 °C, ±0.10 °C at 660 °C
- Axial Uniformity: ±0.05 °C at 50 °C, ±0.35 °C at 420 °C, ±0.5 °C at 660 °C (40 mm)
- Large LCD display with °C/°F selection
- Dual-zone control for excellent temperature uniformity
- Heating Time: 15 minutes (50 °C to 660 °C)
Applications
- Loop calibration of temperature transmitters
- RTD and thermocouple calibration in field environments
- Automated thermal sensor testing and documentation
- Transmitter loop verification and adjustment
- Industrial thermometer verification
- Thermostatic switch testing with auto and manual procedures
Included Accessories
- Insert B with holes for 1/4 in (2), 3/16 in (2), 1/8 in (1), 3/8 in (1) probes
- RS-232 serial cable
- Model 9930 Interface-it control software for test export
- Test leads for transmitter connections
- Users manual and technical manual
Specifications
| Temperature Range | 50 °C to 660 °C |
| Display Accuracy | ±0.35 °C at 50 °C, ±0.35 °C at 420 °C, ±0.5 °C at 660 °C |
| Stability | ±0.03 °C at 50 °C, ±0.04 °C at 420 °C, ±0.05 °C at 660 °C |
| Radial Uniformity | ±0.02 °C at 50 °C, ±0.05 °C at 420 °C, ±0.10 °C at 660 °C |
| Axial Uniformity (40 mm) | ±0.05 °C at 50 °C, ±0.35 °C at 420 °C, ±0.5 °C at 660 °C |
| Immersion Depth | 150 mm (5.9 in) |
| Heating Time | 15 minutes (50 °C to 660 °C) |
| Weight | 7.7 kg (17 lbs) |
Request a quote
Fluke 9144-B-P-156 Field Metrology Well
Response within 24 hours
No obligation
Direct communication