Fluke Calibration 9143-C Dry-Well Field Calibrator
Request a quote
Field metrology well with 33°C to 350°C range, ±0.2°C accuracy, and Insert C configuration for high-volume probe calibration.
Product details
| Model | 9143-C |
| Manufacturer | Fluke Calibration |
| Category | Dry Block Calibrators |
| Availability | Made to order |
Description
Overview
Fluke Calibration's 9143-C is a field metrology well calibrator with temperature range of 33°C to 350°C and display accuracy of ±0.2°C. Designed for high-volume thermocouple and RTD probe calibration, it comes with Insert C (six 0.25-inch holes) for standardized probe testing in field and laboratory environments. This portable unit delivers metrology-grade accuracy while maintaining ease of transport and operation.
Key Features
- Temperature range: 33°C to 350°C (91°F to 662°F)
- Display accuracy: ±0.2°C across full range
- High stability: ±0.02°C at 33°C and 200°C, ±0.03°C at 350°C
- Fast heating: reaches 350°C in 5 minutes from 33°C
- Insert C configuration: six 0.25-inch holes for standardized probe calibration
- Lightweight and portable: 7.3 kg with compact footprint
- Large LCD display with Celsius/Fahrenheit selection
- Excellent axial and radial uniformity for consistent measurements
- Stability indicator shows readiness for calibration
- Easy-to-navigate menus and function keys
- RS-232 interface with Interface-it software
Applications
- High-volume thermocouple and RTD probe calibration
- Industrial temperature sensor verification
- Field and laboratory metrology
- Calibration of standardized probe sets
- Temperature equipment validation
Included Accessories
- Insert C (six 0.25-inch holes)
- RS-232 cable and Interface-it software
Specifications
| Temperature Range | 33°C to 350°C (91°F to 662°F) |
| Display Accuracy | ±0.2°C Full Range |
| Stability at 33°C | ±0.02°C |
| Stability at 200°C | ±0.02°C |
| Stability at 350°C | ±0.03°C |
| Heating Time (33°C to 350°C) | 5 minutes |
| Cooling Time (350°C to 33°C) | 32 minutes |
| Cooling Time (350°C to 100°C) | 14 minutes |
| Immersion Depth | 150 mm (5.9 in) |
| Insert OD | 25.3 mm (1.00 in) |
| Display | LCD, selectable °C or °F |
| Resolution | 0.01° |
| Size | 290 mm H × 185 mm W × 295 mm D |
| Weight | 7.3 kg (16 lb) |
Request a quote
Fluke Calibration 9143-C Dry-Well Field Calibrator
Response within 24 hours
No obligation