BSL Equipment
Home / Calibration Instruments / Temperature Calibrators / Dry Block Calibrators / Fluke Calibration 9142-DW-P Dry-Well Field Calibrator

Fluke Calibration 9142-DW-P Dry-Well Field Calibrator

Fluke Calibration · Model: 9142-DW-P Made to order
Fluke Calibration 9142-DW-P Dry-Well Field Calibrator
Request a quote

Fluke Calibration 9142-DW-P portable dry-well calibrator covers -25°C to 150°C with ± 0.2°C accuracy and built-in two-channel readout.

Model 9142-DW-P
Manufacturer Fluke Calibration
Category Dry Block Calibrators
Availability Made to order

Overview

The Fluke Calibration 9142 is a portable dry-well field metrology well designed for rapid temperature calibration. With the ability to cool to -25°C in 15 minutes and heat to 150°C, it combines speed with the accuracy and stability required for professional calibration work. The unit features a large LCD display and intuitive controls, making it suitable for field technicians who need reliable, traceable calibration capability without carrying additional equipment.

Key Features

  • Temperature range: -25°C to 150°C
  • Accuracy: ± 0.2°C (full range)
  • Stability: ± 0.01°C (full range)
  • Radial uniformity: ± 0.01°C
  • Axial uniformity at 40 mm: ± 0.05°C
  • Built-in two-channel readout for PRT, RTD, thermocouple, and 4-20 mA current
  • Visual and audible stability indicator
  • Onboard non-volatile memory for documentation of up to 20 tests
  • Pre-programmed calibration routines
  • NIST-traceable IEC-17025 NVLAP-accredited calibration certificate

Applications

  • Loop calibration of temperature transmitters
  • Thermocouple calibration or verification
  • RTD or PRT calibrations
  • Testing of thermostatic switches
  • Verification of industrial thermometers
Temperature Range -25°C to 150°C
Accuracy ± 0.2°C (full range)
Stability ± 0.01°C (full range)
Radial Uniformity ± 0.01°C (full range)
Axial Uniformity at 40 mm ± 0.05°C
Immersion Depth 150 mm (6 in)
Cooling Time to -25°C 15 minutes
Heating Time to 150°C 15 minutes
Display Large LCD
Memory Non-volatile memory for up to 20 tests
Temperature Sensor Input PRT, RTD, Thermocouple, 4-20 mA current
Calibration Certification NIST-traceable IEC-17025 NVLAP-accredited