BSL Equipment
Home / Calibration Instruments / Temperature Calibrators / Dry Block Calibrators / Fluke Calibration 9142-C-P-156 Field Metrology Well

Fluke Calibration 9142-C-P-156 Field Metrology Well

Fluke Calibration · Model: 9142-C-P Made to order
Fluke Calibration 9142-C-P-156 Field Metrology Well
Request a quote

Fluke Calibration 9142-C-P-156 combines portable field metrology with built-in process electronics for direct sensor measurement.

Model 9142-C-P
Manufacturer Fluke Calibration
Category Dry Block Calibrators
Availability Made to order

Overview

The Fluke Calibration 9142-C-P-156 Field Metrology Well is a portable temperature calibration instrument enhanced with integrated process electronics. This P-series variant eliminates the need for external measurement devices by providing built-in readout for PRTs, RTDs, thermocouples, and 4-20 mA current loops, with integrated 24 V loop power supply. The combination of metrology-grade temperature control and direct sensor measurement makes it an all-in-one field calibration solution.

The unit maintains the excellent temperature stability and uniformity of the base 9142 design while adding automation capabilities for documentation and data logging. Its compact, portable form factor makes it suitable for demanding field calibration work where efficiency and accuracy are critical.

Key Features

  • Temperature range: -25°C to 150°C
  • Display accuracy: ±0.2°C full range
  • Stability: ±0.01°C full range
  • Axial uniformity at 40 mm: ±0.05°C
  • Radial uniformity: ±0.01°C
  • Loading effect: ±0.006°C full range
  • Hysteresis: 0.025°C
  • Immersion depth: 150 mm (5.9 in)
  • Insert C: six 0.25-inch probe holes
  • Rapid heating: 16 min (23°C to 140°C), 25 min (-25°C to 150°C)
  • Rapid cooling: 15 min (23°C to -25°C)
  • Process electronics: direct measurement of PRTs, RTDs, thermocouples, and 4-20 mA transmitters
  • Built-in 24 V loop power for transmitter operation
  • Two-channel thermometer readout
  • RS-232 and 9930 Interface-it software for automation and documentation
  • Size: 290 × 185 × 295 mm; Weight: 8.16 kg

Applications

  • Temperature transmitter loop calibration with direct current measurement
  • Thermocouple sensor calibration and verification
  • RTD and PRT calibration
  • 4-20 mA current loop transmitter calibration
  • Thermostatic switch testing
  • Automated field calibration documentation
Temperature Range -25°C to 150°C
Display Accuracy ±0.2°C full range
Stability ±0.01°C full range
Radial Uniformity ±0.01°C
Axial Uniformity at 40 mm ±0.05°C
Loading Effect ±0.006°C full range
Hysteresis 0.025°C
Immersion Depth 150 mm (5.9 in)
Resolution 0.01°C
Heating Time to 140°C 16 minutes
Cooling Time to -25°C 15 minutes
Process Electronics Two-channel thermometer readout for PRTs, RTDs, thermocouples, 4-20 mA transmitters
Loop Power Supply 24 V DC
Display LCD, °C or °F selectable
Size 290 × 185 × 295 mm
Weight 8.16 kg (18 lb)
Fluke Calibration 9142-C-P-156 Field Metrology Well

Request a quote

Fluke Calibration 9142-C-P-156 Field Metrology Well

Response within 24 hours
No obligation
Direct communication

By submitting this form, you agree to the processing of your personal data in accordance with our privacy policy.