Fluke Calibration 9142-C-P-156 Field Metrology Well
Fluke Calibration 9142-C-P-156 combines portable field metrology with built-in process electronics for direct sensor measurement.
Product details
| Model | 9142-C-P |
| Manufacturer | Fluke Calibration |
| Category | Dry Block Calibrators |
| Availability | Made to order |
Description
Overview
The Fluke Calibration 9142-C-P-156 Field Metrology Well is a portable temperature calibration instrument enhanced with integrated process electronics. This P-series variant eliminates the need for external measurement devices by providing built-in readout for PRTs, RTDs, thermocouples, and 4-20 mA current loops, with integrated 24 V loop power supply. The combination of metrology-grade temperature control and direct sensor measurement makes it an all-in-one field calibration solution.
The unit maintains the excellent temperature stability and uniformity of the base 9142 design while adding automation capabilities for documentation and data logging. Its compact, portable form factor makes it suitable for demanding field calibration work where efficiency and accuracy are critical.
Key Features
- Temperature range: -25°C to 150°C
- Display accuracy: ±0.2°C full range
- Stability: ±0.01°C full range
- Axial uniformity at 40 mm: ±0.05°C
- Radial uniformity: ±0.01°C
- Loading effect: ±0.006°C full range
- Hysteresis: 0.025°C
- Immersion depth: 150 mm (5.9 in)
- Insert C: six 0.25-inch probe holes
- Rapid heating: 16 min (23°C to 140°C), 25 min (-25°C to 150°C)
- Rapid cooling: 15 min (23°C to -25°C)
- Process electronics: direct measurement of PRTs, RTDs, thermocouples, and 4-20 mA transmitters
- Built-in 24 V loop power for transmitter operation
- Two-channel thermometer readout
- RS-232 and 9930 Interface-it software for automation and documentation
- Size: 290 × 185 × 295 mm; Weight: 8.16 kg
Applications
- Temperature transmitter loop calibration with direct current measurement
- Thermocouple sensor calibration and verification
- RTD and PRT calibration
- 4-20 mA current loop transmitter calibration
- Thermostatic switch testing
- Automated field calibration documentation
Specifications
| Temperature Range | -25°C to 150°C |
| Display Accuracy | ±0.2°C full range |
| Stability | ±0.01°C full range |
| Radial Uniformity | ±0.01°C |
| Axial Uniformity at 40 mm | ±0.05°C |
| Loading Effect | ±0.006°C full range |
| Hysteresis | 0.025°C |
| Immersion Depth | 150 mm (5.9 in) |
| Resolution | 0.01°C |
| Heating Time to 140°C | 16 minutes |
| Cooling Time to -25°C | 15 minutes |
| Process Electronics | Two-channel thermometer readout for PRTs, RTDs, thermocouples, 4-20 mA transmitters |
| Loop Power Supply | 24 V DC |
| Display | LCD, °C or °F selectable |
| Size | 290 × 185 × 295 mm |
| Weight | 8.16 kg (18 lb) |
Request a quote