BSL Equipment
Home / Temperature Measurement / Temperature Probes & Sensors / Fluke Calibration 5627-9-B RTD Probe

Fluke Calibration 5627-9-B RTD Probe

Fluke Calibration · Model: 5627-9-B Made to order
Fluke Calibration 5627-9-B RTD Probe
Request a quote

Fluke Calibration 5627-9-B RTD probe with 100Ω resistance, -200°C to 420°C range, 316 stainless steel sheath, ±0.050°C accuracy, and vibration resistance.

Model 5627-9-B
Manufacturer Fluke Calibration
Category Temperature Probes & Sensors
Availability Made to order

Overview

The Fluke Calibration 5627-9-B is a precision RTD probe that delivers accurate temperature measurement in industrial and laboratory settings. With a 100Ω resistance element and 316 stainless steel sheath, it provides stable performance across -200°C to 420°C and withstands vibration and shock.

Key Features

  • Temperature range: -200°C to 420°C
  • Resistance: 100Ω with temperature coefficient of 0.00385Ω/Ω/°C
  • Accuracy: ±0.050°C at -196°C and 0°C; ±0.051°C at 200°C; ±0.055°C at 420°C
  • Sheath material: 316 stainless steel
  • Minimum bend radius: 3/4 inch (19 mm) for durability
  • Vibration and shock resistant
  • Teflon-insulated leads with nickel-plated copper (22 AWG)
  • Time constant: 4 seconds maximum (63.2% response)
  • Drift rate: ±0.13°C at 0°C after 1000 hours at 400°C

Applications

  • Industrial temperature monitoring and process control
  • Laboratory and calibration testing
  • High-temperature measurement systems
  • Cryogenic and low-temperature applications

Included Accessories

  • Accredited calibration certificate
Resistance 100Ω nominal
Temperature Coefficient 0.00385Ω/Ω/°C nominal
Temperature Range -200°C to 420°C
Accuracy ±0.050°C at -196°C and 0°C; ±0.051°C at 200°C; ±0.055°C at 420°C
Drift Rate ±0.13°C at 0°C after 1000 hours at 400°C
Sheath Material 316 stainless steel
Lead Type Teflon-insulated, nickel-plated copper, 22 AWG
Lead Termination Bare wire
Minimum Bend Radius 19 mm (3/4 in)
Immersion Depth At least 100 mm (4 in) recommended
Time Constant 4 seconds maximum (63.2% response at 3 fps)
Physical Size 9 in length x 3/16 in outer diameter
Fluke Calibration 5627-9-B RTD Probe

Request a quote

Fluke Calibration 5627-9-B RTD Probe

Response within 24 hours
No obligation
Direct communication

By submitting this form, you agree to the processing of your personal data in accordance with our privacy policy.