Fluke 5615-12-L Secondary PRT Probe
Fluke 5615-12-L is a secondary reference platinum resistance thermometer with 100 ohm resistance, 305 mm Inconel 600 sheath, temperature range 196–420°C, and mini spade lug termination.
Product details
| Model | 5615-12-L |
| Manufacturer | Fluke |
| Category | Temperature Probes & Sensors |
| Availability | Made to order |
Description
Overview
The Fluke 5615-12-L is a platinum resistance thermometer (PRT) with an Inconel 600 sheath measuring 305 mm (12 in) in length and 6.35 mm (0.250 in) in diameter. It serves as a secondary reference temperature standard, bridging the gap between top-tier laboratory standards and industrial or secondary lab settings. The sensing element is made from reference-grade platinum wire (99.999% pure) for superior stability and provides accuracy of ±0.012°C at 0.01°C resolution.
The platinum wire is coiled and placed on a mandrel for uniform support that minimizes hysteresis. The four-wire current-potential electrical configuration eliminates lead-wire resistance effects. The Inconel-sheathed probe features a fully supported sensing element in an ultrahigh-purity ceramic case with hermetic glass seal, improving output stability by preventing moisture and contaminant infiltration.
Key Features
- Temperature range: 196°C to 420°C
- Sheath: 305 mm length, 6.35 mm diameter, Inconel 600 material
- Maximum immersion depth: 254 mm (10 in)
- Reference-grade platinum sensing element (99.999% pure)
- 4-wire PTFE insulated lead cable, 22 AWG, 183 cm (72 in) length
- Mini spade lug termination
- ITS-90 calibration coefficients for broad compatibility
Applications
- Secondary reference standard for laboratory calibration
- Industrial temperature measurement and process monitoring
- Secondary laboratory applications requiring high accuracy
Specifications
| Temperature Range | 196°C to 420°C |
| Sheath Length | 305 mm (12 in) |
| Sheath Diameter | 6.35 mm (0.250 in) |
| Sheath Material | Inconel 600 |
| Maximum Immersion Length | 254 mm (10 in) |
| Sensing Element | Reference-grade platinum (99.999% pure) |
| Resistance | 100 Ohm |
| Lead Cable | 4-wire PTFE insulated, 22 AWG |
| Lead Length | 183 cm (72 in) |
| Termination | Mini spade lugs |