Request a quote

Noise Com CNG-EBNO-IF1 is a precision SNR noise generator covering 50-90 MHz and 100-180 MHz bands with ±0.2 dB RSS accuracy for IF testing applications.

Model CNG-EBNO-IF1
Manufacturer Noise Com
Category RF & Microwave Test Equipment
Availability Made to order

Overview

The Noise Com CNG-EBNO-IF1 is a fully automated precision SNR noise generator covering the 50-90 MHz and 100-180 MHz intermediate frequency (IF) bands. It maintains highly accurate carrier-to-noise, carrier-to-interferer, and energy-to-noise density ratios, making it ideal for IF stage testing in military, satellite, and cellular receiver development.

Key Features

  • C/N, C/No, Eb/No, and C/I ratio measurement and generation
  • Frequency coverage: 50-90 MHz and 100-180 MHz
  • 6.25-inch color TFT touch screen with simultaneous readout of signal levels
  • Accuracy: ±0.2 dB RSS (±0.3 dB worst-case)
  • Output power range: -55 dBm to +5 dBm
  • Bit rate entry: 1 bps to 999 Mbps and above
  • True RMS power meter with Gaussian crest factor up to 18 dB

Applications

  • IF stage testing in military and satellite receivers
  • Cellular and mobile receiver development and validation
  • Signal-to-noise ratio characterization
  • General purpose IF band noise generation
Multiple Operating Modes The CNG-EbNo provides five operating modes: carrier-to-noise (C/N), carrier-to-noise density (C/N₀ ), bit energy-to-noise density (Eb/N₀ ), carrier-to-interferer (C/I), and power meter. The instrument can also be used as a precision noise generator.
Custom Configurations The CNG-EbNo is available in a variety of configurations to meet your specific testing needs. Applications include: military communications WCDMA SATCOM NASA TDRSS CableTV HDTV IS-95 CDMA TSMA UMTS GPRS L-band modems Milstar Inmarsat Intelsat General Purpose.
Direct Display of Eb/No, C/N, C/I, or C/No The 6.25” color TFT touch screen provides simultaneous readout of all significant input and output signal levels relating to the chosen operating mode, including carrier-to-noise ratios.
Accuracy of 0.2 dB RSS A special, large-dynamic-range power meter measures both the signal and the noise, which allows the CNG-EbNo to set the desired ration to within ±0.2 dB. Special configurations can provide improved accuracy.
Bit Rate Entry of 1 bps to 999 Mbps and above In bit energy-to-noise density testing (Eb/No ), the instrument automatically calculates noise density based on the user specified bit rate.
Variable Output Power Output power is user-specified and can be set within the range of -55 dBm to +5 dBm.
True RMS Power Meter The digital power meter is custom designed to cover the frequency range of the particular instrument. It can measure signals and noise accurately with Gaussian Noise crest factors up to 18 db.
Direct Testing at both RF and Microwave Frequencies In configurations that cover two separate frequency ranges, measurements can be made directly without the need for special conversion circuitry.
Feature Parameter
Carrier Path Input Power Range: -55 dBm to +5 dBm Maximum Input Power: +21 dBm (with no damage) Output Power Range: -55 dBm to +5 dBm
Noise Path Output Power Range: -55 dBm to +5 dBm Ratio Accuracy: ±0.2 dB RSS, ±0.3 dB WCU Power Meter Range: -55 dBm to +5 dBm Power Meter Accuracy: ±0.5 dB
Primary Power Operating Temperature: 0° to 50° C Dimensions: 17” W x 5.25” H x 17.5” D
CNG-EbNo-IF1 50 to 90 MHz 100 to 180 MHz