Fluke 752A provides 10:1 and 100:1 ratio divider outputs with 0.2 ppm and 0.5 ppm uncertainties, featuring built-in self-calibration without external standards.

Model 752A
Manufacturer Fluke
Category Electrical Calibrators
Availability Made to order

Overview

The Fluke 752A Reference Divider sets the standard for ratio accuracy and ease of use. It offers two divider outputs with 10:1 and 100:1 ratios and output uncertainties of less than 0.2 ppm and 0.5 ppm respectively. The device features a built-in calibration bridge with dynamic resistor matching, allowing self-calibration with only a stable voltage source and a null detector in approximately five minutes without requiring external standards.

Key Features

  • 10:1 and 100:1 divider outputs
  • Output uncertainties of 0.2 ppm and 0.5 ppm respectively
  • Built-in calibration bridge with dynamic resistor matching
  • Self-calibration capability without external standards
  • System switching for ease of use
  • Compensation for long-term changes in divider resistor values

Applications

  • Precision voltage and ratio calibration
  • Reference standard maintenance
  • Ratio verification and accuracy benchmarking
Ratio Ranges  10:1 and 100:1
Ratio Uncertainty Ratio accuracies that apply for a temperature variation of less than ± 1°C from the self-calibration temperature (between 18°C and 28°C) for up to 8 hours following self-calibration: -------- Range 10:0 Input Voltage: 0 to 100V Output Uncertainty: 0.2 ppm Null Accuracy: ±0.5 mV**  -------- Range 100:1 Input Voltage: 0 to 1000V Output Uncertainty: 0.5 ppm  Null Accuracy: ±1.0 mV**
Temperature Coefficient £ ± 1 ppm per °C over range of 18°C to 28°C (typically 0.1 ppm per °C from 15°C to 30°C
Input Resistance 10:1 Ratio: 380 kΩ ± 1% 100:1 Ratio: Divider is 4 MΩ; Driven Guard is 4 MΩ; total is 2 MΩ ± 1%
Maximum Input Voltage 200V for the 10:1 ratio 1100V for the 100:1 ratio
Power Coefficient £ 0.05 ppm of output with 100V applied for 10:1 ratio and £ 0.3 ppm of output with 1000V applied for 100:1 ratio (included in the ratio accuracy specifications)
Temperature 0°C to 50°C, operating; -40°C to 75°C non-operating
Relative Humidity £ 75% to 40°C, £ 45% to 50°C, non-condensing, operating; £ 100% 10-50°C, non-operating
Altitude £ 3050m (10,000 ft) operating;£ 12,220m (40,000 ft) non-operating
Vibration Per MIL-T-28800C, Type III, Class 5, Style E
Safety IEC 348, 2nd edition, 1978; ANSI-C39.5, 1980, CSA 556B, and UL 1244
Calibration Documentation Statement of calibration practices included; performance check optional
Size 19.1 cm H x 22.1 cm W x 60.3 cm L (7.53 in H x 8.69 in W x 23.75 in L)
Weight 8.4 kg (18.5 lbs.)