Fluke 9170-F-R-156 metrology well achieves -45°C to 140°C temperature range with ±0.005°C stability and integrated reference thermometer input for precision calibration.
Product details
| Model | 9170-F-R-156 |
| Manufacturer | Fluke |
| Category | Calibration Instruments |
| Availability | Made to order |
Description
Overview
The Fluke 9170-F-R-156 is a precision metrology well designed for temperature calibration applications. It achieves the lowest temperatures in the Fluke 9170 series, reaching -45°C in normal room conditions and extending to 140°C. With exceptional stability of ±0.005°C over its full temperature range and 160 mm (6.3 in) of immersion depth, this model delivers precise temperature control suitable for pharmaceutical and industrial calibration work.
Key Features
- Temperature range: -45°C to 140°C (-49°F to 284°F) at 23°C ambient
- Display accuracy: ±0.1°C full range
- Stability: ±0.005°C full range
- Axial uniformity: ±0.1°C at -45°C, ±0.02°C at 0°C, ±0.07°C at 140°C
- Radial uniformity: ±0.01°C full range
- Well depth: 160 mm (6.3 in)
- Resolution: 0.001°C
- Built-in reference thermometer input with Tweener technology
- RS-232 serial interface for automated calibration
- 8 temperature set points with ramp and soak programming
- LCD display with user-selectable °C or °F
- Insert F: Aluminum with metric comparison holes (3 mm, 4 mm, 6 mm)
Applications
- Calibration of RTDs, thermocouples, and thermistors
- Pharmaceutical temperature testing and validation
- Industrial temperature sensor calibration
- Precision temperature measurement verification
Included Accessories
- Insert F (aluminum with metric comparison holes)
- Power cord
- RS-232 cable
- Tongs (insert removal tool)
- 9930 Interface-it Software
- Getting Started Guide and User's Guide
- Report of Calibration
Specifications
| Series | 9170 Series |