Fluke 5500A-SC Oscilloscope Calibrator
Request a quote
Fluke 5500A-SC multi-product calibrator with integrated oscilloscope calibration up to 600 MHz for comprehensive electrical verification.
Product details
| Model | 5500A-SC |
| Manufacturer | Fluke |
| Category | Calibration Instruments |
| Availability | Made to order |
Description
Overview
The Fluke 5500A-SC is a comprehensive electrical calibrator with integrated oscilloscope calibration capability. It sources direct and alternating voltage and current with multiple waveforms, simulates power with phase control, and can source resistance, capacitance, and temperature signals from thermocouples and RTDs. The integrated oscilloscope calibration function provides level sine wave, fast edge, time mark, and amplitude signals for oscilloscope calibration up to 600 MHz.
Key Features
- DC voltage range 0 to ±1020V with ±50 ppm accuracy
- DC current range 0 to ±11A with ±0.01% accuracy
- AC voltage 1 mV to 1020V (10 Hz to 500 kHz) with ±0.03% accuracy
- AC current 29 mA to 11A (10 Hz to 10 kHz) with ±0.06% accuracy
- Resistance source 0 to 3.29 MΩ with ±0.009% accuracy
- Capacitance source 0.33 nF to 1.1 mF with ±0.25% accuracy
- Thermocouple simulation -250°C to +2316°C (11 types)
- RTD source -200°C to +630°C (8 types)
- DC and AC power sources up to 11 kW with ±0.08% to ±0.15% accuracy
- Phase control 0 to ±179.99° with ±0.15° accuracy
- Frequency range 0.01 Hz to 2.0 MHz with ±25 ppm accuracy
- Integrated oscilloscope calibration up to 600 MHz
- IEEE-488 (GPIB), RS-232, and optional 5725A interfaces
Applications
- Oscilloscope and ScopeMeter test tool calibration
- Handheld and bench multimeter calibration
- Wattmeter and power analyzer calibration
- Thermocouple and RTD thermometer calibration
- Data logger and strip chart recorder calibration
- Process calibrator and current clamp calibration
- Comprehensive electrical instrument calibration
Specifications
| Features | Parameters |
| Voltage DC | Range: 0 to ±1020V Best 1 Year Specification: ±50 ppm of setting |
| Current DC | Range: 0 to ±11A Best 1 Year Specification: ±0.01% of setting |
| Resistance | Range: 0 to 3.29.999 MΩ Best 1 Year Specification: ±0.009% of setting |
| Voltage AC | Range: 1 mV to 1020V 10 Hz to 500 kHz, Sine Best 1 Year Specification: ±0.03% of setting |
| Current AC | Range: 29 m A to 11A 10 Hz to 10 kHz, Sine Best 1 Year Specification: ±0.06% of setting |
| Capacitance | Range: 0.33 nF to 1.1 mF Best 1 Year Specification: ±0.25% of setting |
| RTD Source | Range: -200°C to +630°C (8 types) Best 1 Year Specification: ±0.03°C of setting |
| DC Power | Range: 109 mW to 11 kW Best 1 Year Specification: ± 0.08% of setting |
| AC Power | Range: 109 mW to 11 kW 45-65 Hz, PF=1 Best 1 Year Specification: ± 0.15% of setting |
| Phase | Range: 0 to ± 179.99° Best 1 Year Specification: ±0.15° |
| Frequency | Range: 0.01 Hz to 2.0 MHz Best 1 Year Specification: ±25 ppm of setting |
| Warm up Time | Twice the time since last warmed up, to a maximum of 30 minutes |
| Settling Time | Less than 5 seconds for all functions and ranges except as noted |
| Standard Interfaces | IEEE-488 (GPIB), RS-232, 5725A (5500A only) |
| Temperature Performance | Operating: 0°C to 50°C Calibration: (tcal): 15°C to 35°C Storage: -20°C to 70°C |
| Temperature Coefficient | Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C |
| Relative Humidity | Operating: < 80% to 30°C, < 70% to 40°C, < 40% to 50°C Storage: < 95%, non-condensing |
| Altitude | Operating: 3,050m (10,000 ft) maximum Non-operating: 12,200m (40,000 ft) maximum |
| Safety | Designed to comply with: IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92 |
| Analog Low Isolation | 20V |
| EMC | Designed to comply with FCC Rules Part 15 |
| Line Power | Line Voltage (selectable): 100V, 120V, 220V, 240V Line Frequency: 47 to 63 Hz Line Voltage Variation: ± 10% about line voltage setting |
| Power Consumption | 300 VA |
| Calibration Documentation | 17025 accredited report of calibration included |
| Dimensions | Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit Width: 43.2 cm (17 in), standard rack width Depth: 47.3 cm (18.6 in) overall |
| Weight (without options) | 20 kg (44 lbs.) |
| Absolute Uncertainty Definition | 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated. |
| Specification Confidence Interval | > 99% |
| 5500A-SC300 | 300 MHz Oscilloscope Calibration Option |
| 5500A-SC600 | 600 MHz Oscilloscope Calibration Option |
Request a quote