Description
Overview
The Fluke 5080A is a high-compliance multi-product calibrator designed to provide accurate calibration of analog meters, panel meters, digital multimeters, and other measurement instruments. With maximum voltage compliance of 800 mA for AC/DC voltage and 50 V for AC/DC current, the 5080A offers the versatility needed for a wide range of industrial and laboratory applications.
Key Features
- High voltage and current compliance (600 mA DC, 800 mA AC for voltage measurement; 50 V DC/AC for current measurement)
- DC voltage range 0 to ±1020 V with ±0.01% best specification accuracy
- DC current range 0 to ±20.5 A with ±0.05% best specification accuracy
- Resistance measurement 0 to 190 MΩ with ±0.025% accuracy
- AC voltage 1 mV to 1020 V (45 Hz to 1 kHz) with ±0.1% specification accuracy
- AC current 29 mA to 20.5 A (45 Hz to 1 kHz) with ±0.1% accuracy
- Phase measurement 0 to ±179.99° with ±0.25° accuracy
- RS-232 and Ethernet interfaces for remote operation and data logging
- Protection circuitry for safe operation
Applications
- Calibration of analog meters and gauges
- Testing of 3.5 and 4.5 digit digital multimeters
- Panel meter and watt meter calibration
- Clamp meter calibration (with optional coil accessory)
- Megohm meter calibration (optional)
- Oscilloscope calibration to 200 MHz (optional)
Specifications
| Features | Parameters |
| Voltage DC | Range: 0 to ±1020V Best 1 Year Specification: ±0.01% of setting |
| Current DC | Range: 0 to ±20.5A Best 1 Year Specification: ±0.05% of setting |
| Resistance | Range: 0 to 190 MO 19 values in x1 and x1.9 Best 1 Year Specification: ±0.025% of setting |
| Voltage AC | Range: 1 mV to 1020V 45Hz to 1 kHz, Sine Best 1 Year Specification: ±0.1% of setting |
| Current AC | Range: 29 mA to 20.5A 45 Hz to 1 kHz, Sine Best 1 Year Specification: ±0.1% of setting |
| DC Power | Range: 10.9 mW to 20.5 kW Best 1 Year Specification: ±0.11% of setting |
| AC Power | Range: 10.9 mW to 20.5 kW 45-65 Hz, PF=1 Best 1 Year Specification: ±0.36% of setting |
| Phase | Range: 0 to ± 179.99° 45 Hz to 1 kHz Best 1 Year Specification: ±0.25° of setting |
| Frequency | Range: 45 Hz to 1 kHz Best 1 Year Specification: ±50 ppm of setting |
| Warm up Time | Twice the time since last warmed up, to a maximum of 30 minutes |
| Settling Time | Less than 5 seconds for all functions and ranges except as noted |
| Standard Interfaces | RS-232 and Ethernet |
| Temperature Performance | Operating: 0°C to 50°C Calibration (tcal):15°C to 35°C Storage:-20°C to +70°C |
| Temperature Coefficient | Temperature coefficient for temperatures outside tcal ±5 °C is 10 % of the stated specification per °C for temperatures in the range of 0 °C to 35 °C. Above 35 °C, the temperature coefficient is 20 % of the stated specification per °C. |
| Relative Humidity | Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C Storage: <95%, non-condensing |
| Altitude | Operating: 2,000m (6,500 ft) maximum Non-operating: 12,200m (40,000 ft) maximum |
| Safety | Meets EN. 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1: 2004 |
| Analog Low Isolation | 20V |
| EMC | Meets EN613261:2006. |
| Line Power | Line Voltage (selectable): 100V, 120V, 220V, 240V Line Frequency: 47 to 63 Hz Line Voltage Variation: + 10% about line voltage setting |
| Power Consumption | 600 VA |
| Calibration Documentation | Report of calibration with data included; 17025 accredited report optional |
| Dimensions | Height: 19.3 cm (7.6 in) Width: 43.2 cm (17 in), 44.3 cm (17.5 in) including handles Depth: 53.8 cm (21.2 in) |
| Weight (without options) | 22 kg (48 lbs.) |
| Absolute Uncertainty Definition | The specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. It is not necessary to add anything to determine the total specification for the temperature range indicated. |
| Specification Confidence Interval | 99% |
| 5080A | 600 mA800 mA50 V44 V |